Page 25 of 43
6.3
Instrument Setup
6.3.1
Display Unit
0.30
[%]
H2/N2
Unit -> Vol%
T90 Response
------------
------------
Unit -> ppm
------------
T90 Response
------------
Diagnosis
Instr. Setup
Output Setup
Expert Setup
Figure 6.4: changing the displayed unit for gas concentrations
The first item of the instrument setup menu allows selection of the display unit. Pressing <ENTER>
alternates the unit between ppm and Vol.%. Quit the menu with the <UP> key. The value of parameter
P56 (see Section 6.5.1) sets the number of digits after the decimals point between 1 and 4 when the
display unit is set to Vol.%. The resolution in ppm is always 1 ppm. All internal calculations of the
FTC320 are done in ppm.Values retrieved through the RS-232 interface will always be in ppm with
1ppm resolution.
6.3.2
Response Time Setup
Unit -> Vol%
T90 Response
------------
------------
T90 Response
set: 1.00000
new: 1.00000
[s]
ESC
/OK
0.30
[%]
H2/N2
Diagnosis
Instr. Setup
Output Setup
Expert Setup
Figure 6.5: T90 response time setup
In the menu “T90 Response“ the response time of the exponential filter can be adjusted. The filter
reduces the influence of fast variations of the raw signal (signal noise smoothing). The numerical value
for the T90 response time is given in the unit of seconds. A range between 0s and 100s is permitted for
this value, reasonable values lie between 0.5s to 10s. You can change the numerical value by navigat-
ing through the digit position using the <RIGHT> key and changing it using the <UP> key. Confirm by
pressing the <UP> key until ”OK” is selected. The changed value can be discarded by selecting ”ESC”
instead. Setting “T90 Response“ to 0.0 turns the exponential filter off.
The true response time is influenced by the gas exchange time which depends on the
pneumatic installation and the flow rate of measuring gas. A gas flow of 80l/h leads to a
gas exchange time of under 0.5s measured from the gas inlet of the device.
Note that the response time in your process depends greatly on the upstream volume
before the FTC320. Shorter and/or thinner tubing will benefit response times.