• Relay 2 (Rel 2) (open/closed)
• Common Relay (Rel 3) (open/closed)
• Current of current loop
• Voltage for Analog Out 1
• Voltage for Analog Out 2
The digital input “DIN“ is “low“ for voltages below 4.6V and “high“ above 11.4 V. The output “Current
Loop“ can be set to currents between 0 and 22mA. The analog outputs 1 and 2 can be set to supply
a voltage between 0V and 10V.
Note!
All test signals are permanently on until leaving the “I/O Test“ menu.
It is
the responsibility of the expert to assure that the I/O test does not interfere with the
subsequent connected systems and processes in an unintended way.
File name:
FTC300 Manual 1.09KD140704CWI1V05.pdf