• “R 1 Mode L/H“: signalling from low to high. When the measured concentration increases and
reaches the threshold, the alarm is active until the measured concentration decreases again
below the threshold minus a hysteresis value. The hysteresis value is specified in percent of
threshold (can be set to 0%).
• “R 1 Mode H/L“: signalling from high to low. When the measured concentration decreases and
reaches the threshold, the alarm is active until the measured concentration increases again
above the threshold plus a hysteresis value. The hysteresis is specified in percent of threshold.
• “R 1 Mode OFF“: Limit value alarm is deactivated, the relay is not used.
8.3.2
Relay 1 Threshold
A1 A2 Er
0.30
[%]
H2/N2
Operation
Calibration
Diagnosis
Setup
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Relay 1
Relay 2
------------
Common Relay
R1 Mode ***
Failsafe
Cal. active
ESC/OK
R1 Mode L/H
R1 Threshold
R1 Hyst.
ESC/OK
R1 Threshold
set: 100.000
new: 100.000
[ppm]
ESC
/OK
Figure 8.8: Relay 1 threshold configuration
In the “R1 Threshold“ menu the threshold value is set. The value is set in the currently used measur-
ing unit (ppm in this case). The threshold is independent of the measuring range.
You can navigate to the digit position you wish to change using the <RIGHT> key. A digit’s value can
be changed using the <UP> key. To confirm your entry, press the <RIGHT> key several times until
“OK“ is highlighted and press <ENTER>. Selection of “ESC“ will discard made changes.
8.3.3
Relay 1 Hysteresis
A1 A2 Er
0.30
[%]
H2/N2
Operation
Calibration
Diagnosis
Setup
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Relay 1
Relay 2
------------
Common Relay
R1 Mode ***
Failsafe
Cal. active
ESC/OK
R1 Mode L/H
R1 Threshold
R1 Hyst.
ESC/OK
R1 Hyst.
set: 1.00000
new: 1.00000
[%]
ESC
/OK
Figure 8.9: Relay 1 hysteresis configuration
File name:
FTC300 Manual 1.09KD140704CWI1V05.pdf
Содержание FTC300
Страница 1: ...FTC300 Gas analysis using thermal conductivity measurement Operating Manual ...
Страница 55: ...Page 54 of 58 This page is intentionally left blank File name FTC300 Manual 1 09KD140704CWI1V05 pdf ...
Страница 58: ...Page 57 of 58 Chapter 12 Appendix Dimensional Drawing File name FTC300 Manual 1 09KD140704CWI1V05 pdf ...