Chapter 6
Calibration
6.1
General Information on Calibration
For devices operated in “safety mode“, the operator code is required to access the calibration menu.
By default the operator code is set to 111.000.
We recommend a calibration, resp. a check of the calibration if one of the following criteria is met:
• After bringing the device into service
• On a regular cycle, depending on the precision aimed for. To find out the appropriate time
between calibrations, we recommend to begin with a more frequent check of the calibration.
The time between calibrations can range between:
–
several month for a measuring task in the Vol.% range
–
days to weeks for a measuring task in the sub-Vol.% range
–
directly before every measurement if highest accuracy is needed
• When the situation of the measurement regarding pressure, temperature or gas flow changes
The goal of the calibration is that the measured concentration is in agreement with the given test gas
concentration. To obtain this, two calibration parameters that correspond to the offset/zero and the
gain/span of a linear equation are available. A two-point calibration requires two test gases. Both
calibration parameters, offset and gain, are adjusted. The concentration of the test gases does not
have to meet the beginning and the end of the measuring range, a difference of
±
10% is permitted.
The menu sequence is designed such that prior to a gain always an offset calibration
has to be done first. Usually a single point calibration determining a new offset value is
sufficient to obtain a good calibration.
In the case of one point calibration, a test gas of any concentration in the measuring
range is feasible. For two point calibrations it is preferable to use the gas concentrations
at the lower and upper end of your measuring range.
6.1.1
Calibration gas purities and flooding time
Messkonzept uses gases with the following purities for calibration:
File name:
FTC300 Manual 1.09KD140704CWI1V05.pdf