8.3.5
Relay 1 active/frozen during calibration
A1 A2 Er
0.30
[%]
H2/N2
Operation
Calibration
Diagnosis
Setup
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Relay 1
Relay 2
------------
Common Relay
Alarm matrix
Failsafe
Cal. active
ESC/OK
Alarm matrix
Failsafe
Cal. frozen
ESC/OK
Figure 8.11: Setup of relay behaviour during calibration
The relays can be configured as “Calibration active“ or “Calibration frozen“. With “Calibration frozen“
the relay does not respond to any changes of the measured concentration during the calibration
procedure and stays in the state it was in when the calibration began. With “Calibration active“ the
relay acts according to its current settings during calibration. Press <ENTER> to switch between
“Calibration frozen“ and “Calibration active“. The change is made effective by pressing <RIGHT> two
times to mark “OK“ and confirming with <ENTER>.