which a parameter’s value can be modified.
Setting certain parameters to improper values can cause falty measurement results,
malfunctions or even permanent physical destruction of the device!
8.5.2
Access Modes
Parameter
Access Modes
Reset Funct.
I/O Test
Operat. Code
Expert Code
Operat. Mode
Operat. Code
set: 111.000
new: 111.000
ESC
/OK
Operat. Code
Expert Code
Operat. Mode
Operat. Code
Expert Code
Operat. Mode
Expert Code
set: 222.000
new: 222.000
ESC
/OK
Operat. Mode
set: 1.00000
new: 1.00000
ESC
/OK
Figure 8.21: Access Mode menu
In this menu the passcodes for the “Operator Mode“ and the “Expert Mode“ can be changed. Further,
the device’s mode can be switched between “Safety Mode“ and “Operation Mode“. For the “Operation
Mode“ set “1.00000“ in the menu “Mode Set“, for the safe mode enter “3.00000“. In the safe mode,
every change of settings requires manual input of the Operator Code.
8.5.3
Reset Functions
Parameter
Access Modes
Reset Funct.
I/O Test
Restart only
Factory Set.
Default Set.
Save Calibr.
Figure 8.22: Reset Functions menu
This menu provides three reset functions and the possibility to save the current calibration parameters
to the Factory Settings:
• “Restart only“: Restart of the software
• “Factory Set.“: Resets all parameters to the factory set values
File name:
FTC300 Manual 1.09KD140704CWI1V05.pdf