Electrical Connectors and Ground
1: Analog Input 1
2: Analog Output 2
3: Analog Reference
4: Digital Input (24V)
5: Ground pin 4
6: Analog Output 1
7: Analog Input 2
1: 0 (4) - 20mA,
is
2: 0 (4) - 20mA,
isolated -
3: Ground
RS232
4: TxD
5: RxD
1: Relay 1 - A
2: Power Supply-GND
4: Power 24V
3: Relay 1 - B
7: Relay 3 - A
6: Relay 2 - B
5: Relay 2 - A
8: Relay 3 - B
A
B
C
Figure 3.3: Electrical connector pin assignments of the three connectors on the FTC300
The FTC300 has three three plug connectors as shown in Figure 3.3. Further information on the
function of each contact is given in Table 3.1. The cables (712, IP67) with molded connector plug
and a length of two meters (five meters available on request) are part of the purchased parts pack-
age. The cables have open ends. The cross-section of the conductors in cable A and C is 0,14mm
2
,
for cable B 0,25mm
2
. Cable A is shipped with devices set up for analog output.
The protection class of the device is only effective with all cables attached. In case cable
A is not used, connector plug A has to be closed with an end fitting.
File name:
FTC300 Manual 1.09KD140704CWI1V05.pdf
Содержание FTC300
Страница 1: ...FTC300 Gas analysis using thermal conductivity measurement Operating Manual ...
Страница 55: ...Page 54 of 58 This page is intentionally left blank File name FTC300 Manual 1 09KD140704CWI1V05 pdf ...
Страница 58: ...Page 57 of 58 Chapter 12 Appendix Dimensional Drawing File name FTC300 Manual 1 09KD140704CWI1V05 pdf ...