Page 37 of 58
8.3.6
Relay 2
A1 A2 Er
0.30
[%]
H2/N2
Operation
Calibration
Diagnosis
Setup
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Relay 1
Relay 2
------------
Common Relay
R2 Mode ***
Failsafe
Cal. active
ESC/OK
R2 Mode L/H
R2 Threshold
R2 Hyst.
ESC/OK
R2 Mode H/L
R2 Threshold
R2 Hyst.
ESC/OK
Aux. L/H
R2 Threshold
R2 Hyst.
ESC/OK
Aux H/L
R2 Threshold
R2 Hyst.
ESC/OK
Maintenance
R2 Threshold
R2 Hyst.
ESC/OK
R2 Mode OFF
------------
------------
ESC/OK
Figure 8.12: Setup of the additional Maintenance mode option of Relay 2
“Relay 2“ provides all functionalities of Relay 1, see description above. In addition Relay 2 permits
the external signaling of the “Maintenance“ alarm or monitoring an external auxiliary signal with the
menu “Aux“. The ”Maintenance”-Alarm is always indicated by the yellow flashing LED on the front
panel. The FTC300 is still functioning but should be put into maintenance in the near future. ”Aux
H/L” and ”Aux L/H” can be used to monitor an external signal that is routed to the analog input of
the device. All other settings for relay 2 can be made analogously as for relay (see Sections 8.3.1
to 8.3.5).