In the “R1 Hysteresis“ menu the hysteresis value for the reset of the relais can be configured. The
value is given in percent of the threshold value.
You can navigate to the digit position you wish to change using the <RIGHT> key. A digit’s value can
be changed using the <UP> key. To confirm your entry, press the <RIGHT> key several times until
“OK“ is highlighted and press <ENTER>. Selection of “ESC“ will discard made changes.
8.3.4
Relay 1 failsafe / Not failsafe
A1 A2 Er
0.30
[%]
H2/N2
Operation
Calibration
Diagnosis
Setup
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Relay 1
Relay 2
------------
Common Relay
Alarm matrix
Not failsafe
Cal. active
ESC/OK
Alarm matrix
Failsafe
Cal. active
ESC/OK
Figure 8.10: Toggling between relay "Failsafe" and "Not failsafe"-mode
The threshold alarms can be configured as “Failsafe“ or “Not Failsafe“. “Failsafe“ means that the
relay is live with no error pending. Hence, the alarm is indicated if the threshold is reached or if the
closed-circuit contact is interrupted. “Not Failsafe“ means that the relay is open with no error pending
and closes when an error appears. Press <ENTER> to switch between “Failsafe“ and “Not Failsafe“.
The change is made effective by pressing <RIGHT> three times to mark “OK“ and confirmed with
<ENTER>.
File name:
FTC300 Manual 1.09KD140704CWI1V05.pdf
Содержание FTC300
Страница 1: ...FTC300 Gas analysis using thermal conductivity measurement Operating Manual ...
Страница 55: ...Page 54 of 58 This page is intentionally left blank File name FTC300 Manual 1 09KD140704CWI1V05 pdf ...
Страница 58: ...Page 57 of 58 Chapter 12 Appendix Dimensional Drawing File name FTC300 Manual 1 09KD140704CWI1V05 pdf ...