![MES SQA-V Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/mes/sqa-v/sqa-v_service-manual_1775133007.webp)
and
SQA-V
Service Manual Dec_2019
7
Rear Panel
Power inlet with fuse-holder (2 fuses: 250V, 2A)
RCA Video connector
RS232 communication port
Specimen Testing Supplies
Measurement capillary:
Disposable, plastic, positive displacement testing
capillary. (Testing depth of capillary section: 300-micron; Cuvette section:
10 mm).
Standard laboratory slide:
20 micron depth, 22 x 22 mm cover-slip.
Operating System
Control:
Keypad
Analysis Time:
Normal Test – 75 seconds; High Sensitivity/Postvasectomy
Test – 5 minutes (9 minutes in version 2.43).
Software:
Resides on flash memory and drives all man-machine interface
functions, runs algorithms for test measurements (according to WHO
guidelines), and operates visual and automated screens. System can be
upgraded from a PC CD-ROM.
Sample Testing Temperature:
Calibrated for room temperature only
20-25° C / 68-77° F. Motility results will be impacted by heating the specimen.
Motility channel input signal:
Analog, up to 5V.
Spectrophotometer channel input signal:
Modulated (1 kHz) analog, up to
5V.
Quality Control
Internal:
Electronic Self-Test and Auto-Calibration.
External:
Counts can be run on the SQA-V automated and visualization
systems using commercially available assayed control material:
"QwikCheck-
beads™
" (product of Medical Electronic Systems) or Stabilized Sperm as a
non-assayed control in version 2.45 and higher.
PC Compatibility
Minimum requirements for V-Sperm software
PC:
1 GHz processor, Pentium 3
RAM:
256 MB
AGP-video display card
with at least 16 MB of RAM memory
Video color
: At least 16 bit (65,535)
CD ROM drive
200 MB free hard disk space
for image capturing
Video resolution:
Minimum 640 x 480
Operating system compatibility:
Windows XP, VISTA, Windows 7,
(Windows 10 pending)
Ports:
One serial; two USB ports
Monitor:
15" color
Required Software
V-Sperm software for SQA-V system and I-button set-up, archive and
CONTROL reporting
also permits user to benefit from many features such as
real time visualization interface between PC and SQA-V visualization system,
data analysis, video/picture capture and archives test records.
Excel/Word for exporting files and clips