and
SQA-V
Service Manual Dec_2019
13
#
Parameter
Software version
2.45,2.46, 2.48, 2.49
SQA-V Value
Pass
Fail
1.
Ref 1
150 – 350 mV
2.
LED Cur 1
5 – 25 mA
3.
Amplitude
50 – 100 mV
4.
Zero Level
500 - 525
5.
Ref 2
2500 – 3500 mV
6.
LED Cur 2
10 – 32 mA
7.
CONC. 1
0 – 1 M/ml
8.
CONC. 2
50-150 M/ml
9.
CONC. 3
300-600 M/ml
10.
Count (Internal
Data, Item #12)
26 - 36
SECTION VI: Getting Started / Set-Up
Please refer to the relevant SQA-V User Guide for start-up and set-up instructions (based
upon the software version of the SQA-V).
NOTE:
Turn off the
power supply
to the SQA-V
and disconnect
the power
supply cable
from the back
of the device
before opening
the SQA-V.
Only a
qualified, MES
certified
distributor is
authorized to
open the SQA-
V without
voiding the
warranty.
SECTION VII: Troubleshooting
Overview: The following sections describe how to troubleshoot and repair potential
problems that may occur with the SQA-V. Theses section will discuss how to provide
support for three types of issues:
Technical, Self-test, Clinical
PLEASE NOTE:
Only a qualified MES distributor who has been trained to perform technical
support is authorized to open the SQA-V.
Opening SQA-V device without authorization may cause to the calibration
corruption and will VOID THE WARRANTY.
The electro-optical board should NEVER be touched when the SQA-V is
opened – it will cause damage to the SQA-V calibration.
SECTION VIII: Technical Support/Part Replacement
1.
Opening the SQA-V
Turn off the main switch located on
the rear panel and disconnect
the SQA-V from the mains.
Using a Philips screwdriver, unscrew
all the screws on the rear panel.
Opening SQA-V rear panel