High-End Pressure Controller
CPC8000
62
Operating Instructions - CPC8000
7.3 Control Tab
The [Control] tab within the Setup [
] menu contains settings that limit or control display elements
associated with the pressure output from the instrument’s regulator. Control limits, stable window confgu-
ration, and the rate setpoint can be found in this section.
7.3.1 Max and Minimum Limits
The [Maximum Limit] and [Minimum Limit] buttons under the Setup [
] / [Control] screen provide a
place to limit the setpoint that can be chosen in the main screen. These limits can only be set within the
range of the active sensor. When the CPC8000 is in Autorange the limits can only be set within the range
of the primary transducer which, by convention, will have the widest range. The minimum limit must be
lower than the maximum limit. The operator cannot enter setpoints and thereby not control to pressures
outside of these limits.
Figure 7.3.1 - Maximum Limit and Minimum Limit Buttons