High-End Pressure Controller
CPC8000
Operating Instructions - CPC8000
79
Figure 7.5.2.5.6b - Programs
Table - 7.5.2.5.6 Sequence Commands
Command
Function (data selection)
DELAY
Delays for time = 1 to 3600 seconds (Numerical Entry)
MODE
Sets the control mode (Measure, Control or Vent)
PTYPE
Sets the pressure type (Gauge or Absolute)
RSETPT
Sets the rate setpoint in current units (Numerical Entry)
RUNITS
Sets the rate denominator time unit (min or sec)
SENSOR
Sets the active sensor (1 , 2 or 3)
SEQSTART
Starts the sequence from the beginning (None)
SETPOINT
Sets the control setpoint for the instrument (Numerical Entry)
SETPOINT%
Sets the control setpoint in % of current range (Numerical Entry)
WAIT
Waits for a manual input or stable condition (Stable or Input)