High-End Pressure Controller
CPC8000
46
Operating Instructions - CPC8000
6.2.14.1 Measure Mode
In Measure mode, the instrument measures the pressure at the sensor connected to the Measure/
Control port. The Measure mode is activated by pressing the [Measure] button.
Figure 6.2.14.1a - Measure Mode
Measure
Mode
If the pressure is beyond the permissible measurement range, the pressure reading is displayed in red
instead of white fgures. As soon as a pressure of approximately 110% full scale is reached, an integrated
safety relief valve opens and releases the pressure into the inside of the instrument. The regulator is not
active in Measure mode.
The setpoint value can be input while in the Measure mode then Control mode can be activated by press-
ing the [Control] key and the CPC8000 will start to control to the setpoint value.