Menu functions
7.3.7 Balance
For appliances of the size 55 and above, application-specific correction of the heat output dis-
tribution (balance) between the upper and lower heating groups is possible. The adjustment
range is from –50 % to +50 %.
upper heat output
upper heat output
lower heat output
-20%
lower heat output
+30%
SETUP
Balance
-20
%
2/2
SETUP
Balance
+30
%
2/2
Fig. 25
Heat output distribution (example): The –20 % (left) setting causes the lower heating
groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower
Fig. 25
Fig. 25
Heat output distribution (example): The –20 % (left) setting causes the lower heating
Heat output distribution (example): The –20 % (left) setting causes the lower heating
heating groups to emit 30 % more heat than the upper ones. The 0 % setting restores the
groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower
groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower
default heat output distribution.
heating groups to emit 30 % more heat than the upper ones. The 0 % setting restores the
heating groups to emit 30 % more heat than the upper ones. The 0 % setting restores the
7.3.8 Remote control
In the setup entry
Remote control
, you can set whether the
appliance should be controlled via remote control and if
so, in which mode. These settings are available:
►
Off
►
Read Only
►
Write
►
Write+Alarm
When the appliance is in remote control mode, the
symbol appears in the temperature display. In the
symbol appears in the temperature display. In the
settings
Write
and
Write+Alarm
, the appliance cannot
be controlled at the ControlCOCKPIT until the remote
control has been switched off (setting Off) or set to
Read
Only
.
In order to use the remote control function, program-
ming skills and special libraries are required.
Setup
Balance
Remote Control
Off
Gateway 192.168.5 .1
+30 %
2/2
23
.2°C
TEMP
Set 180
.0°C