28
Operation and control
5.4.5 Operation with digital backwards counter with target time setting,
adjustable from 1 minute to 99 days (Timer)
In timer operation, you can adjust the time the appliance runs at the set value. The appliance
has to be in manual operating mode for this.
1. Press the activation key to the left of
the timer display. The timer display is
activated.
30.0
%rh
50.0
%CO
2
Programm 12
Fr 20.10.2010
20:31
min
000°C
max
000°C
.5°C
100
ICH 500
MENU
30.0
%rh
50.0
%CO
2
Programm 12
Fr 20.10.2010
20:31
min
000°C
max
000°C
MENU
TIMER
-
Ende 9:00 23.11.
--
h
-
m
2. Turn the turn control until the desired
duration is displayed – in this example 4
hours 30 minutes. The approximate end
time is shown beneath, in a smaller font.
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroffen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344
.4
444
4
.
TEMP
Set
°C
°C
FLAP
40
%
TIMER
h
d
10
02
End
Sept.
29
22 24
FAN
%
0
%
100
ALARM
max
444
.4°C
auto
+/-
0.0
K
min
444
.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
TIMER
End
0
13:30 23.11.
04
m
h
3
Up to a duration of 23 hours 59 minutes, the time is displayed in hh:mm (hours:minutes)
format. For 24 hours and more, the format dd:hh (days:hours) is used. The maximum
duration adjustable is 99 days 00 hours.
3. Press the confirmation key to confirm.
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroffen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344
.4
444
4
.
TEMP
Set
°C
°C
FLAP
40
%
TIMER
h
d
10
02
End
Sept.
29
22 24
FAN
%
0
%
100
ALARM
max
444
.4°C
auto
+/-
0.0
K
min
444
.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
The display now shows the remaining time
in a large font and the approximate end time
in a smaller font beneath. The status display
shows „Timer active“.
TIMER
30
m
04
h
Ende 13:30 23.11.
Timer active
12.Sept.2012 13:44
4. Now, as described under 5.4.2 , set the individual values for temperature, air flap position
etc. which you want the appliance to operate at. For universal ovens UN
PLUS
/UF
PLUS
PLUS
PLUS
and
incubators IN
PLUS
/IF
PLUS
PLUS
PLUS
, the set values can be changed while the timer elapses. The changes
are effective immediately. For sterilisers SN
PLUS
/SF
PLUS
PLUS
PLUS
, parameters cannot be changed while
the timer elapses.
For universal ovens UN
PLUS
/UF
PLUS
PLUS
PLUS
and incubators IN
PLUS
/IF
PLUS
PLUS
PLUS
, you can choose if the timer
should run setpoint-dependent or not in the
Setup
– this determines if the timer should
not start until a tolerance band around the set temperature is reached or if it should start
right after activation (see page 44). If the timer runs setpoint-dependent, this is
indicated by the
symbol in the timer display.
When the timer has elapsed, the display shows 00h:00m. All func-
tions (heating etc.) are switched off. If a fan had been active, it will
keep on running for a short safety period. In addition, an acoustic
alarm sounds, which can be turned off by pressing the confirma-
tion key.
TIMER
00
m
00
h
End 13:30 23.11.
Содержание IF PLUS
Страница 59: ......