Malfunctions, warning and error messages
6.3 Power failure
In case of a power failure, the appliance operates as follows:
In manual mode
After power supply has been restored, operation is continued with the parameters set. The
time and duration of the power failure are documented in the log memory.
In timer or programme mode
In timer or programme mode
In case of an interruption of the power supply of less than 60 minutes, the current pro-
gramme is continued from the point at which it was interrupted. For interruptions of the
power supply longer than this, all appliance functions (heating, fan etc.) are switched off and
the air flap is opened.
In timer or programme mode of sterilisers
In timer or programme mode of sterilisers
After power supply has been restored, the timer or programme always starts again from the
beginning.
In remote control mode:
The previous values are restored. If a programme has been initiated via remote control, it is
continued.