9018116
41 / 54
9.9
Service level
Code display
Meaning
View parameter, see page 41.
Vent rinse pipe, see page 41.
Vent detergent/disinfectant pipe, see page
41.
Reset the counter for replacing the partial
desalination cartridge, see page 42.
9.9.1
View parameters
1. Switch to authorisation level 1
Service level (10000)
, see chapter “Change au-
2. Select the entry 1
–1.
3. Confirm the selection.
The first parameter is displayed.
4. Scroll through and view the parameters using the cleaning programme keys.
The service level can be departed with the
on/off key
.
9.9.2
Ventilating the pipes
The ventilation of the detergent/disinfectant or rinse pipes must be performed if air is
sucked in from the dosing units. This occurs if a canister is completely emptied dur-
ing operation, or if one of the suction lances is not threaded through to the base of
the canister.
1. Switch to authorisation level 1
Service level (10001)
, see page 40.
2.Select entry 1-3 for venting the detergent/disinfectant line, and entry 1-2 for the
rinse pipe.
3. Confirm the selection.
The respective pipes are being ventilated, and the remaining time in seconds is
displayed. If necessary, repeat ventilation.
Ventilation can be cancelled with the
on/off key
.