38
.2 Set the rotary selector switch to the
2P
position.
.3 Press and release the TEST button.
The instrument will perform pre-measurement check, the status of which will be indicated on the display.
The two-terminal resistance reading will be displayed
Note:
The test voltage used to make the two-terminal resistance reading is a.c. and may not be suitable for continuity testing
according to some local regulations.
9.4
Making a measurement – Three terminal resistance measurement
.1
Connect the instrument as below.
I
CAT IV
300V
12-15V
1.2A
Electrode
under test
Potential
stake
Current
stake
.2
Set the rotary selector switch to the
3P
position.
.3 Press and release the TEST button. The instrument will perform pre-measurement check, the status of which will be indicated on
the display.
The three-terminal resistance reading will be displayed