41
10. SETUP OPTIONS
The Setup options allow the MFT to be configured to best suit the type of testing to which it will be used.
To enter SETUP, right (secondary) range knob to
SETUP. Set the left (primary) range knob to any function other than OFF.
He display will show VER and the software version number. It will then change to the first message in the list below:
Message
Function
Options
Factory setting
RST
Restore factory settings
NO / YES
NO
INS*1
Insulation limit alarm – Buzzer
sounds if result is higher than
limit set
0.5,1,2,3,4,5,7,
10,50,100,500M
W
1 M
W
LOC
Insulation test lock.
ON / OFF
ON
bUZ
Continuity limit alarm – Buzzer
sounds if result is below than
limit set
0.5,1,2,5,10,
50,100
W
2
W
ISC*2
Continuity test current
15mA / 200mA
200mA
REV
Auto reverse continuity test
ON / OFF
OFF
looP
Loop test lead compensation
0 – 0.3ohms
0.07
W
LAS
Loop test AUTO start
ON/OFF
OFF
L-PE 2Hi
Enable/Disable high current
loop test
ON/OFF
ON
L-PE 2Lo
Enable/Disable 2-wire low
current loop test
ON/OFF
ON
RAS
RCD AUTO start
ON/OFF
OFF
RRA*3
RCD Ramp test
Nor = Normal, FST=Fast
Nor / FST
Nor
UL
Touch voltage limit
25V / 50V / 60V
50V
N –L
Reverse Polarity Loop test
enable
ON= Loop test allowed
OFF= Loop test stopped
ON/OFF
OFF
OFF
Auto switch OFF in minutes
2m / 10m / off
20minutes
bAt
Alkaline or NiMH selection
1.5V or 1.2V
Depending on instrument
StR
Store mode
IN = Internal
Bt = Bluetooth only
IN / bT
IN
bt
Bluetooth pairing
bt1, bt2, bt3, bt4, bt5
bt1
< >
Searching for pair
*1INS is not available on MFT1815
*2ISC is not available on MFT1815
*3RRA is only available on MFT1800 series
To scroll through the options, press the
button. Each option will be is displayed in sequence.
To change the setting of each function, for example, INS limit alarm from 1M
W
to 2M
W
, use the right hand TEST and LOCK keys (also
marked with UP/DOWN arrows).
Changing an option will set the LOCK symbol and warning triangle flashing.
To save the change press the left LOCK button
To exit SETUP, turn the right range knob away from
All settings can be restored to the factory defaults by setting RST to YES. Saving this setting will reset all options to default. The RST will
then set back to NO.