background image

1.  Set the selector switch to mA.
2.  Connect the test leads.
3.  Press and hold down the TEST button for approximately 2 seconds.
4.  Select either ac or dc mA using the

key.

5.  After a short settling time, the reading will be displayed

automatically.

To view the frequency of the a.c. mA being measured press the

key. The frequency will be displayed in the range 16Hz-

460Hz. To view a.c. mA press the

key again. For inputs

less than 10mA frequency is not displayed.

Using the Megger SP1 Switched Probe
Operation: 
The  Megger SP1 is an accessory for designated
MEGGER installation test instruments. When fitted in the
specially designed connector, in place of the existing ‘Low’ lead,
the SP1 acts as a remote test button to operate the instrument
and as a ‘Low’ probe. This simplifies instrument control and two-
handed probing. The SP1 is suitable for use with Megger
insulation test instruments up to 1kV output test voltage.
Safety: Meets the safety requirements for double insulation to
IEC1010-2-031 (1995), EN61010-2-031 (1995), IEC1010-1
(1995), EN61010-1 (1995) Category III*, 300V phase to earth and
500V phase to phase. The probe is fitted with an internal, non-
replaceable fuse, to protect the user should the probe be used
accidentally in conjunction with a test lead in the low terminal.
* Relates to transient overvoltage likely to be found in fixed

installation wiring.

Do not use the probe if any part of it is damaged.

Battery Replacement
When the low battery symbol

appears, the cells are nearly

exhausted and should be replaced as soon as possible. Use
Alkaline cells IEC LR6 (AA) or NiCd rechargeable. To install or
replace the cells, disconnect the test leads, switch the
instrument to OFF and loosen the captive screws on the rear of
the battery compartment. Remove the cover and disconnect
the battery holder from the battery leads. Ensure that the
replacement cells are fitted with the correct polarity in
accordance with the label in the battery holder. Reconnect the
battery holder to the battery leads. Replace and re-secure the
battery compartment cover. Remove the cells if the instrument
is not going to be used for an extended period of time.

Fuse Checking and Replacement
To check the instrument fuse, switch to an insulation range and
press the TEST button. The symbol

will appear if the fuse

is ruptured. To replace the fuse, disconnect the test leads,
switch the instrument OFF and loosen the captive screws
holding the battery compartment cover in place. Remove the
cover and replace the fuse. Replace and re-secure the battery
compartment cover.

~

Содержание BMM2000 Series

Страница 1: ...ave no broken or cracked insulation Ensure that hands remain behind guards of probes clips when testing U K Safety Authorities recommend the use of fused test leads when measuring voltage on high ener...

Страница 2: ...Regulations IEC364 and HD384 since each range has a 1mA minimum test current at the minimum pass values of insulation specified in these documents The 100V range is ideal for testing telecommunication...

Страница 3: ...is activated by pressing the key The backlight will remain illuminated for approx one minute before automatically switching off to conserve battery life alternatively the key can be re pressed Insula...

Страница 4: ...paid to the terminals Also attempt to reduce any leakage that may give erroneous results on the item under test Polarisation Index Testing Polarisation Index PI is the term applied to the Dielectric A...

Страница 5: ...ation is retained in non volatile memory and so will be remembered when the instrument is switched off 1 Select the Continuity range 2 Short the test leads across a known good conductor using prods 3...

Страница 6: ...ent The diode symbol will appear if the voltage developed across the terminals is within semi conductor junction limits These features together with the small test current and wide measurement range 0...

Страница 7: ...icate the zero has been adjusted 1 Select the uF range 2 Disconnect the test leads from the circuit under test 3 When the reading has stabilised press the TEST button The zero offset symbol will appea...

Страница 8: ...he user should the probe be used accidentally in conjunction with a test lead in the low terminal Relates to transient overvoltage likely to be found in fixed installation wiring Do not use the probe...

Страница 9: ...digits 0 8 per G 100V 20G 2 2 digits 2 0 per G 50V 10G 2 2 digits 4 0 per G Accuracy BMM2000 Range Full Scale Accuracy 1000V 20G 2 2digits 0 2 per G 500V 10G 2 2 digits 0 4 per G 250V 5G 2 2 digits 0...

Страница 10: ...460 Hz 5 7 digits 10mV to 1999mV a c 16 460 Hz 5 5 digits d c milliVolts zeroing Up to 9 9mV Input resistance 3M Capacitance BMM2080 Measuring Range 0 1nF to 9 99uF Accuracy 3 2 digits 0 2nF uF zeroin...

Страница 11: ...imit Min Indicated Limit Min Indicated Reading Reading 0 10 0 14 2 00 2 12 0 20 0 25 3 00 3 16 0 30 0 35 4 00 4 20 0 40 0 46 5 00 5 24 0 50 0 56 10 00 10 8 0 60 0 66 20 00 21 2 0 70 0 77 30 00 31 6 0...

Страница 12: ...sec 500V insulation tests 4000 5 sec 250V insulation tests 2700 5 sec continuity tests 4700 5 sec k tests ENVIRONMENTAL CONDITIONS Operating range 5 to 40 C Operating humidity 90 RH at 40 C max Stora...

Страница 13: ...contact Megger Limited or Megger Archcliffe Road Valley Forge Corporate Center Dover 2621 Van Buren Avenue Kent CT17 9EN Norristown PA 19403 England U S A Tel 44 0 1304 502243 Tel 1 610 676 8500 Fax 4...

Страница 14: ...rporate Center 2621 Van Buren Avenue Norristown PA 19403 Tel 1 610 676 8500 U S A Fax 1 610 676 8610 Megger Z A Du Buisson de la Couldre 23 rue Eug ne Henaff T 33 1 30 16 08 90 78190 TRAPPES France F...

Отзывы: