Zeroing of d.c. mV (no a.c. mV zero facility)
To zero the d.c. mV range, short the leads together in the
d.c. mV position, wait for the reading to settle and then press
the TEST button. Up to 9,9mV can be zeroed on the d.c. mV
range. The symbol

will appear to indicate the zero has

been adjusted.
1.  Select the d.c. mV range.
2.  Short the test leads together.
3.  When the reading has stabilised, press the TEST button.

The zero offset symbol

will appear.

4.  To release the zero offset press the TEST button again.

To view the frequency of the a.c. mV being measured press the

key. The frequency will be displayed in the range 16Hz-

460Hz. To view a.c. mV press the

key again. For inputs

less than 10mV frequency is not displayed.

Capacitance Tests (uF) (BMM2080 only)
The measured capacitance is indicated on the display.
1.  Set the selector switch to uF.
2.  Connect the test leads to the circuit under test.
3.  After a short settling time, the reading will be displayed

automatically.

Zeroing of uF
To zero the uF range, disconnect the leads, wait for the reading
to settle and then press the TEST button. Up to 10,0nF can be
zeroed on the uF range. The symbol

will appear to

indicate the zero has been adjusted.
1.  Select the uF range.
2.  Disconnect the test leads from the circuit under test.
3.  When the reading has stabilised, press the TEST button.

The zero offset symbol

will appear.

4.  To release the zero offset press the TEST button again.

The range is suitable for the testing of discrete components
and short low interference level signal lines. If electrolytic
capacitors are being tested then the red lead should be
connected to +ve of the capacitor. This range is not suitable for
checking capacitance of signal lines which are subject to high
levels of a.c. interference.

When the test is started --- will show on the display, if there is
excessive noise this symbol will remain or flash indicating that
there is too much noise for a result to be reached.

Milliamps Tests (mA) (Not BMM2000ESD)
Because of the low source impedance associated with current
measuring this test has an added feature ensuring that when
the range is first entered the default voltmeter is visible. Testing
will be inhibited if more than 25V is present at the terminals. To
start testing the TEST button should be pressed and held down
for approximately 2s to activate the mA range. Once activated,
the TEST button no longer needs to be used and the measured
value will be displayed automatically. To switch the display
between a.c. and d.c. press the

key. 

~

Содержание BMM2000 Series

Страница 1: ...ave no broken or cracked insulation Ensure that hands remain behind guards of probes clips when testing U K Safety Authorities recommend the use of fused test leads when measuring voltage on high ener...

Страница 2: ...Regulations IEC364 and HD384 since each range has a 1mA minimum test current at the minimum pass values of insulation specified in these documents The 100V range is ideal for testing telecommunication...

Страница 3: ...is activated by pressing the key The backlight will remain illuminated for approx one minute before automatically switching off to conserve battery life alternatively the key can be re pressed Insula...

Страница 4: ...paid to the terminals Also attempt to reduce any leakage that may give erroneous results on the item under test Polarisation Index Testing Polarisation Index PI is the term applied to the Dielectric A...

Страница 5: ...ation is retained in non volatile memory and so will be remembered when the instrument is switched off 1 Select the Continuity range 2 Short the test leads across a known good conductor using prods 3...

Страница 6: ...ent The diode symbol will appear if the voltage developed across the terminals is within semi conductor junction limits These features together with the small test current and wide measurement range 0...

Страница 7: ...icate the zero has been adjusted 1 Select the uF range 2 Disconnect the test leads from the circuit under test 3 When the reading has stabilised press the TEST button The zero offset symbol will appea...

Страница 8: ...he user should the probe be used accidentally in conjunction with a test lead in the low terminal Relates to transient overvoltage likely to be found in fixed installation wiring Do not use the probe...

Страница 9: ...digits 0 8 per G 100V 20G 2 2 digits 2 0 per G 50V 10G 2 2 digits 4 0 per G Accuracy BMM2000 Range Full Scale Accuracy 1000V 20G 2 2digits 0 2 per G 500V 10G 2 2 digits 0 4 per G 250V 5G 2 2 digits 0...

Страница 10: ...460 Hz 5 7 digits 10mV to 1999mV a c 16 460 Hz 5 5 digits d c milliVolts zeroing Up to 9 9mV Input resistance 3M Capacitance BMM2080 Measuring Range 0 1nF to 9 99uF Accuracy 3 2 digits 0 2nF uF zeroin...

Страница 11: ...imit Min Indicated Limit Min Indicated Reading Reading 0 10 0 14 2 00 2 12 0 20 0 25 3 00 3 16 0 30 0 35 4 00 4 20 0 40 0 46 5 00 5 24 0 50 0 56 10 00 10 8 0 60 0 66 20 00 21 2 0 70 0 77 30 00 31 6 0...

Страница 12: ...sec 500V insulation tests 4000 5 sec 250V insulation tests 2700 5 sec continuity tests 4700 5 sec k tests ENVIRONMENTAL CONDITIONS Operating range 5 to 40 C Operating humidity 90 RH at 40 C max Stora...

Страница 13: ...contact Megger Limited or Megger Archcliffe Road Valley Forge Corporate Center Dover 2621 Van Buren Avenue Kent CT17 9EN Norristown PA 19403 England U S A Tel 44 0 1304 502243 Tel 1 610 676 8500 Fax 4...

Страница 14: ...rporate Center 2621 Van Buren Avenue Norristown PA 19403 Tel 1 610 676 8500 U S A Fax 1 610 676 8610 Megger Z A Du Buisson de la Couldre 23 rue Eug ne Henaff T 33 1 30 16 08 90 78190 TRAPPES France F...

Отзывы: