background image

Complies with the following parts of EN61557, Electrical safety in
low voltage systems up to 1000V a.c. and 1500V d.c. –
Equipment for testing, measuring or monitoring of protective
measures:-

Part 1 – 

General requirements

Part 2 – 

Insulation resistance

Part 4 –

Resistance of earth connection and equi-
potential bonding

FUSE 

500mA (F) 500V, 32x 6mm Ceramic HBC 10kA minimum.

E.M.C.

The instruments meet EN 61326-1.

POWER SUPPLY

Battery Type:

6x1,5V Alkaline cells IEC LR6 type
or 1.2V NiCd re-chargeable cells.

Battery Life (typical):

2100 5-sec 1kV insulation tests
3200 5-sec 500V insulation tests
4000 5-sec 250V insulation tests
2700 5-sec continuity tests
4700 5-sec k

tests

ENVIRONMENTAL CONDITIONS

Operating range:

-5 to +40°C

Operating humidity:

90% RH at 40°C max.

Storage temperature range: -25 to +65°C
Calibration Temperature:

+20ºC

Maximum altitude:

2000 m

Dust and water protection: IP54
Temperature coefficient:

<0,1% per °C 

WEIGHT:

742g

DIMENSIONS:

110mm x 220mm x 45mm

CLEANING:

Wipe with a clean cloth dampened
with soapy water or Isopropyl
Alcohol(IPA)

ACCESSORIES

Supplied: Part 

Number

Test lead set 

6220-437

Test-&-carry case 

6420-123

Optional:
Fused lead set, FPK8 

6111-218

Switch Test Probe SP1 

6220-606

Test Record Cards (Pack of 20) 

6111-216

Publications:
‘A Stitch in Time’

AVTM21-P8B

‘Testing Electrical Installations’

6231-605

Содержание BMM2000 Series

Страница 1: ...ave no broken or cracked insulation Ensure that hands remain behind guards of probes clips when testing U K Safety Authorities recommend the use of fused test leads when measuring voltage on high ener...

Страница 2: ...Regulations IEC364 and HD384 since each range has a 1mA minimum test current at the minimum pass values of insulation specified in these documents The 100V range is ideal for testing telecommunication...

Страница 3: ...is activated by pressing the key The backlight will remain illuminated for approx one minute before automatically switching off to conserve battery life alternatively the key can be re pressed Insula...

Страница 4: ...paid to the terminals Also attempt to reduce any leakage that may give erroneous results on the item under test Polarisation Index Testing Polarisation Index PI is the term applied to the Dielectric A...

Страница 5: ...ation is retained in non volatile memory and so will be remembered when the instrument is switched off 1 Select the Continuity range 2 Short the test leads across a known good conductor using prods 3...

Страница 6: ...ent The diode symbol will appear if the voltage developed across the terminals is within semi conductor junction limits These features together with the small test current and wide measurement range 0...

Страница 7: ...icate the zero has been adjusted 1 Select the uF range 2 Disconnect the test leads from the circuit under test 3 When the reading has stabilised press the TEST button The zero offset symbol will appea...

Страница 8: ...he user should the probe be used accidentally in conjunction with a test lead in the low terminal Relates to transient overvoltage likely to be found in fixed installation wiring Do not use the probe...

Страница 9: ...digits 0 8 per G 100V 20G 2 2 digits 2 0 per G 50V 10G 2 2 digits 4 0 per G Accuracy BMM2000 Range Full Scale Accuracy 1000V 20G 2 2digits 0 2 per G 500V 10G 2 2 digits 0 4 per G 250V 5G 2 2 digits 0...

Страница 10: ...460 Hz 5 7 digits 10mV to 1999mV a c 16 460 Hz 5 5 digits d c milliVolts zeroing Up to 9 9mV Input resistance 3M Capacitance BMM2080 Measuring Range 0 1nF to 9 99uF Accuracy 3 2 digits 0 2nF uF zeroin...

Страница 11: ...imit Min Indicated Limit Min Indicated Reading Reading 0 10 0 14 2 00 2 12 0 20 0 25 3 00 3 16 0 30 0 35 4 00 4 20 0 40 0 46 5 00 5 24 0 50 0 56 10 00 10 8 0 60 0 66 20 00 21 2 0 70 0 77 30 00 31 6 0...

Страница 12: ...sec 500V insulation tests 4000 5 sec 250V insulation tests 2700 5 sec continuity tests 4700 5 sec k tests ENVIRONMENTAL CONDITIONS Operating range 5 to 40 C Operating humidity 90 RH at 40 C max Stora...

Страница 13: ...contact Megger Limited or Megger Archcliffe Road Valley Forge Corporate Center Dover 2621 Van Buren Avenue Kent CT17 9EN Norristown PA 19403 England U S A Tel 44 0 1304 502243 Tel 1 610 676 8500 Fax 4...

Страница 14: ...rporate Center 2621 Van Buren Avenue Norristown PA 19403 Tel 1 610 676 8500 U S A Fax 1 610 676 8610 Megger Z A Du Buisson de la Couldre 23 rue Eug ne Henaff T 33 1 30 16 08 90 78190 TRAPPES France F...

Отзывы: