All indicators on the control panel will illuminate.
The status panel will display
ANGIOJET ULTRA
while the Console performs a self-test.
The drawer will open, indicating a successful self-
test. The status panel will display the next step
(Figure 5).
1
Sterile technician:
Remove the catheter
and sufficient tubing for ease of use
from the sterile package and inspect for
damage. Hand the tray with the rest of
the Thrombectomy Set to the nonsterile
technician for installation into the Console
(Figure 6).
NOTE: If a clamp is used to keep the catheter
in the sterile field, ensure that the
clamp does not deform the tubing.
2
Nonsterile technician:
Remove the pump
(not the piston head) from the tray and
insert the pump into the Console. Ensure
that the waste tubing aligns with the roller
pump (Figure 7).
3
Remove the cap from the Thrombectomy
Set bag spike and insert the spike into the
saline bag.
4
Push the drawer button to close
the Console drawer (Figure 8).
8
Roller Pump
Waste Tubing
7.2 Load Pump
Figure 6
Figure 5
Figure 8
Delivery Tubing
Figure 7