Alarms and error messages indicate improper Ultra System preparation or
abnormal component operation.
The Console status panel will display alarm resolution messages and
prompts (Example shown in Figure 16). In most instances, alarms will either
adjust themselves or prompt the user to repeat a step. Follow the prompts
displayed on the status panel for alarm resolution.
The Console allows multiple attempts to correct most alarm conditions.
If the alarm persists beyond the third attempt, it is most likely that the
Thrombectomy Set has a defect
and the Console will prompt the user to
replace the Thrombectomy Set (Figure 17).
A system failure will result in a system error (wrench icon lights red, shown
in Figure 18). Turn off the power, then restart before contacting MEDRAD
Technical Service for further instructions.
It is helpful to make a note of each system error title and number to report
to MEDRAD Technical Service.
Refer to your Limited Warranty and Disclaimer and/or Certificate of
Extended Warranty, if applicable, for information on servicing the AngioJet
Ultra Console. MEDRAD recommends annual inspection and calibration.
There are no user serviceable parts inside the Console. The Console cabinet
panels should only be opened by trained service personnel.
7.5 Alarms and Error
Messages
7.6 Ultra System Errors
Figure 18 - Example
Figure 16 - Example of alarm
REPLACE THROMBECTOMY SET
Figure 17 - Example
7.7 Maintenance