Alarms and error messages indicate improper Ultra System preparation or 

abnormal component operation.

The Console status panel will display alarm resolution messages and 

prompts (Example shown in Figure 16). In most instances, alarms will either 

adjust themselves or prompt the user to repeat a step. Follow the prompts 

displayed on the status panel for alarm resolution.

The Console allows multiple attempts to correct most alarm conditions. 

If the alarm persists beyond the third attempt, it is most likely that the 

Thrombectomy Set has a defect

 

and the Console will prompt the user to 

replace the Thrombectomy Set (Figure 17).

A system failure will result in a system error (wrench icon lights red, shown 

in Figure 18). Turn off the power, then restart before contacting MEDRAD 

Technical Service for further instructions. 
It is helpful to make a note of each system error title and number to report 

to MEDRAD Technical Service.

Refer to your Limited Warranty and Disclaimer and/or Certificate of 

Extended Warranty, if applicable, for information on servicing the AngioJet 

Ultra Console. MEDRAD recommends annual inspection and calibration. 

There are no user serviceable parts inside the Console. The Console cabinet 

panels should only be opened by trained service personnel.

7.5 Alarms and Error 

Messages

7.6 Ultra System Errors

Figure 18 - Example

Figure 16 - Example of alarm

REPLACE THROMBECTOMY SET

Figure 17 - Example 

7.7 Maintenance

12

Содержание AngioJet Ultra

Страница 1: ...System Console Operations Manual...

Страница 2: ...S and PRECAUTIONS 3 5 ADVERSE EVENTS 4 6 CONSOLE SPECIFICATIONS 5 7 CLINICIAN USE INFORMATION 6 7 1 Prepare Console 7 7 2 Load Pump 8 7 3 Prime the Catheter 9 7 4 Ultra System Dismantling 11 7 5 Alarm...

Страница 3: ...saline volume and Ultra System malfunction warnings The Console is activated by pressing a foot pedal B The AngioJet Ultra Thrombectomy Set Thrombectomy Set Several models available The single use dis...

Страница 4: ...any of the Console safety features If the catheter is removed from the patient and or is inoperative the waste tubing lumen guide catheter and sheath should be flushed with sterile heparinized solutio...

Страница 5: ...netic interference with other devices when in use Do not place Console near sensitive equipment when operating Equipment not suitable for use in the presence of flammable anesthetic mixture with air o...

Страница 6: ...n IPX8 against ingress of liquid Mode of electrical operation Continuous Transport storage temperature 25 C to 57 C Transport storage humidity 10 to 100 Transport storage atmospheric 500hPa to 1060hPa...

Страница 7: ...ctions on what to do if error messages are displayed Figure 3 Control Panel with Labels Time Display During catheter priming the time display counts down to zero During the procedure the time display...

Страница 8: ...NOTE The Ultra System is designed to be interactive The status panel will provide prompts to guide the technician through set up as well as provide error resolution when necessary Preparation of the...

Страница 9: ...Set to the nonsterile technician for installation into the Console Figure 6 NOTE If a clamp is used to keep the catheter in the sterile field ensure that the clamp does not deform the tubing 2 Nonste...

Страница 10: ...pedal within easy access of the physician Choose a location that will minimize accidental activation The status panel displays PRIME indicating that the initial pump prime was successful Figure 10 Cat...

Страница 11: ...isplays READY and green icon is illuminated Figure 14 The AngioJet Ultra System is now ready to use During operation the infused volume will be displayed on the status panel the green icon remains lit...

Страница 12: ...Press POWER button to deactivate control panel Pump drawer will automatically retract when powering down 5 Unplug the power cord from the wall outlet and coil the cord on the hooks in the power cord a...

Страница 13: ...Figure 17 A system failure will result in a system error wrench icon lights red shown in Figure 18 Turn off the power then restart before contacting MEDRAD Technical Service for further instructions...

Страница 14: ...ing the operator to install the pump into the Console Over Pressure Alarm A recoverable condition occurring when an abnormally high pressure has been detected Replacement of the Thrombectomy Set may b...

Страница 15: ...ator proof type CF equipment 8 IPX 1 Protection against dripping water 9 IPX 8 Protection against the effects of permanent immersion in water 10 Alternating current 11 Equipotentiality 12 Fuse 13 Foot...

Страница 16: ...nd or saphenous vein grafts See Information For Use manuals for specific and complete prescribing information 2006 2009 MEDRAD INC All rights reserved All information drawings and ideas herein constit...

Отзывы: