Follow proper precautions for the handling of infectious waste. Reuse of
Thrombectomy Set is prohibited due to risk of contamination by blood
waste products.
After use, disassemble the components as follows:
1
Push the drawer button to open the drawer.
2
Carefully remove Thrombectomy Set from Console.
3
Unhook the saline supply and collection bag and dispose together with
the Thrombectomy Set.
4
Press
POWER
button to deactivate control panel. Pump drawer will
automatically retract when powering down.
5
Unplug the power cord from the wall outlet and coil the cord on the
hooks in the power cord and foot pedal recess. Place the foot pedal into
the storage bracket at the bottom of the recess.
6
Clean the Console surfaces thoroughly with a standard, mild germicidal
cleaning agent. Do not clean with harsh detergent or chemical agents.
NOTE:
Check bar code window for saline build-up,
clean with long cotton-tipped swab and
water, if necessary (Figure 15).
11
7.4 Ultra System
Dismantling
Figure 15