background image

Status Panel:

  

 

Provides instructions, procedure status, and alarm resolution 

strategies. 

Power Button:

 Press to activate the control panel. 

Alarm Reset Button:

 Press to clear or override an alarm.

  

Catheter Button:

 Displays catheter model installed in the AngioJet 

Ultra Console. Press this button to see a 3-second display of catheter 

model.

 Scroll Buttons: 

These buttons are active only when there is 

more text to be displayed than can fit on the status panel 

screen. Press the 

DOWN

 button to advance one screen 

forward; press the 

UP

 button to return to the previous screen.

Counter Reset Button:

 Press this button to return the time 

display and infused volume to zero.
 

NOTE:   You cannot use the COUNTER RESET Button to override the time needed 

for priming the catheter.

NOTE:   The Ultra System is designed to be interactive. The status panel will 

provide prompts to guide the technician through set-up as well as 

provide error resolution when necessary. 

Preparation of the Ultra System requires the assistance of a sterile and a 

nonsterile technician. The catheter is used within the defined sterile field, 

the Console and pump are operated outside the sterile field. The following 

directions are for the nonsterile technician except where otherwise noted.

1

 

   Plug in the Console and ensure  

the main power circuit breaker 
switch is turned 

ON

 (Figure 4).

2

   Heparinize a bag of sterile, room-temperature 

saline at a suggested rate of 5000 units per 
liter of saline and mix contents (a 1.0 liter bag 
is recommended, but is not included with the 
Ultra System). Hang the saline bag on the saline 
bag hook at the top of the Console.  
 

3

   Press the 

POWER

 button on the control panel. 

7

Figure 4

7.1 Prepare

 

Console

 

Main Power Circuit 

Breaker Switch

Содержание AngioJet Ultra

Страница 1: ...System Console Operations Manual...

Страница 2: ...S and PRECAUTIONS 3 5 ADVERSE EVENTS 4 6 CONSOLE SPECIFICATIONS 5 7 CLINICIAN USE INFORMATION 6 7 1 Prepare Console 7 7 2 Load Pump 8 7 3 Prime the Catheter 9 7 4 Ultra System Dismantling 11 7 5 Alarm...

Страница 3: ...saline volume and Ultra System malfunction warnings The Console is activated by pressing a foot pedal B The AngioJet Ultra Thrombectomy Set Thrombectomy Set Several models available The single use dis...

Страница 4: ...any of the Console safety features If the catheter is removed from the patient and or is inoperative the waste tubing lumen guide catheter and sheath should be flushed with sterile heparinized solutio...

Страница 5: ...netic interference with other devices when in use Do not place Console near sensitive equipment when operating Equipment not suitable for use in the presence of flammable anesthetic mixture with air o...

Страница 6: ...n IPX8 against ingress of liquid Mode of electrical operation Continuous Transport storage temperature 25 C to 57 C Transport storage humidity 10 to 100 Transport storage atmospheric 500hPa to 1060hPa...

Страница 7: ...ctions on what to do if error messages are displayed Figure 3 Control Panel with Labels Time Display During catheter priming the time display counts down to zero During the procedure the time display...

Страница 8: ...NOTE The Ultra System is designed to be interactive The status panel will provide prompts to guide the technician through set up as well as provide error resolution when necessary Preparation of the...

Страница 9: ...Set to the nonsterile technician for installation into the Console Figure 6 NOTE If a clamp is used to keep the catheter in the sterile field ensure that the clamp does not deform the tubing 2 Nonste...

Страница 10: ...pedal within easy access of the physician Choose a location that will minimize accidental activation The status panel displays PRIME indicating that the initial pump prime was successful Figure 10 Cat...

Страница 11: ...isplays READY and green icon is illuminated Figure 14 The AngioJet Ultra System is now ready to use During operation the infused volume will be displayed on the status panel the green icon remains lit...

Страница 12: ...Press POWER button to deactivate control panel Pump drawer will automatically retract when powering down 5 Unplug the power cord from the wall outlet and coil the cord on the hooks in the power cord a...

Страница 13: ...Figure 17 A system failure will result in a system error wrench icon lights red shown in Figure 18 Turn off the power then restart before contacting MEDRAD Technical Service for further instructions...

Страница 14: ...ing the operator to install the pump into the Console Over Pressure Alarm A recoverable condition occurring when an abnormally high pressure has been detected Replacement of the Thrombectomy Set may b...

Страница 15: ...ator proof type CF equipment 8 IPX 1 Protection against dripping water 9 IPX 8 Protection against the effects of permanent immersion in water 10 Alternating current 11 Equipotentiality 12 Fuse 13 Foot...

Страница 16: ...nd or saphenous vein grafts See Information For Use manuals for specific and complete prescribing information 2006 2009 MEDRAD INC All rights reserved All information drawings and ideas herein constit...

Отзывы: