A bag of sterile, heparinized saline (not included) supplies the pump with 

saline through the saline delivery tubing. The pump pressurizes the saline.

 

The

 

Thrombectomy Set uses this pressurized, high-velocity saline to create 

a low-pressure zone at the catheter tip. This allows the catheter to break up 

and remove thrombus. The waste tubing transports the thrombus debris 

from the catheter to the collection bag for ultimate disposal. 

The Console is intended for use only in conjunction with an AngioJet Ultra 

Thrombectomy Set. Refer to the individual Thrombectomy Set

 Information 

for Use

 manual for specific clinical applications.

Refer to the individual Thrombectomy Set

 Information for Use

 manual for 

specific contraindications.

•  The AngioJet Ultra System should be used only by operators who have 

received appropriate training on its installation and use.

•  Use the AngioJet Ultra Console only with an AngioJet Ultra 

Thrombectomy Set. This Console will not operate with a previous  

model pump set and catheter.

•  Do not attempt to bypass any of the Console safety features.
•  If the catheter is removed from the patient and/or is inoperative, the 

waste tubing lumen, guide catheter, and sheath should be flushed with 

sterile, heparinized solution to avoid thrombus formation and maintain 

lumen patency. Reprime the catheter by submerging the tip in sterile, 

heparinized solution and operating it for at least 20 seconds before 

reintroduction to the patient.

•  Physician discretion with regard to the use of heparin is advised.
•  Refer to the individual AngioJet Ultra Thrombectomy Set 

Information 

for Use 

manual for specific warnings and precautions.

•  Do not move the collection bag during catheter operation as this may 

cause a collection bag error.

•  Monitor thrombotic debris/fluid flow exiting the catheter through 

the waste tubing during use. If blood is not visible during console 
activation, the catheter may be occlusive within the vessel or the 
outflow lumen may be blocked. 

•  Ensure adequate patient anticoagulation to prevent thrombus 

formation in outflow lumen. 

•  Refer to individual Thrombectomy Set 

Instructions for Use 

manual for 

specific instructions regarding heparinization of the Thrombectomy Set.

3

2. INDICATIONS AND USAGE

3. CONTRAINDICATIONS

4. WARNINGS AND 

PRECAUTIONS

Содержание AngioJet Ultra

Страница 1: ...System Console Operations Manual...

Страница 2: ...S and PRECAUTIONS 3 5 ADVERSE EVENTS 4 6 CONSOLE SPECIFICATIONS 5 7 CLINICIAN USE INFORMATION 6 7 1 Prepare Console 7 7 2 Load Pump 8 7 3 Prime the Catheter 9 7 4 Ultra System Dismantling 11 7 5 Alarm...

Страница 3: ...saline volume and Ultra System malfunction warnings The Console is activated by pressing a foot pedal B The AngioJet Ultra Thrombectomy Set Thrombectomy Set Several models available The single use dis...

Страница 4: ...any of the Console safety features If the catheter is removed from the patient and or is inoperative the waste tubing lumen guide catheter and sheath should be flushed with sterile heparinized solutio...

Страница 5: ...netic interference with other devices when in use Do not place Console near sensitive equipment when operating Equipment not suitable for use in the presence of flammable anesthetic mixture with air o...

Страница 6: ...n IPX8 against ingress of liquid Mode of electrical operation Continuous Transport storage temperature 25 C to 57 C Transport storage humidity 10 to 100 Transport storage atmospheric 500hPa to 1060hPa...

Страница 7: ...ctions on what to do if error messages are displayed Figure 3 Control Panel with Labels Time Display During catheter priming the time display counts down to zero During the procedure the time display...

Страница 8: ...NOTE The Ultra System is designed to be interactive The status panel will provide prompts to guide the technician through set up as well as provide error resolution when necessary Preparation of the...

Страница 9: ...Set to the nonsterile technician for installation into the Console Figure 6 NOTE If a clamp is used to keep the catheter in the sterile field ensure that the clamp does not deform the tubing 2 Nonste...

Страница 10: ...pedal within easy access of the physician Choose a location that will minimize accidental activation The status panel displays PRIME indicating that the initial pump prime was successful Figure 10 Cat...

Страница 11: ...isplays READY and green icon is illuminated Figure 14 The AngioJet Ultra System is now ready to use During operation the infused volume will be displayed on the status panel the green icon remains lit...

Страница 12: ...Press POWER button to deactivate control panel Pump drawer will automatically retract when powering down 5 Unplug the power cord from the wall outlet and coil the cord on the hooks in the power cord a...

Страница 13: ...Figure 17 A system failure will result in a system error wrench icon lights red shown in Figure 18 Turn off the power then restart before contacting MEDRAD Technical Service for further instructions...

Страница 14: ...ing the operator to install the pump into the Console Over Pressure Alarm A recoverable condition occurring when an abnormally high pressure has been detected Replacement of the Thrombectomy Set may b...

Страница 15: ...ator proof type CF equipment 8 IPX 1 Protection against dripping water 9 IPX 8 Protection against the effects of permanent immersion in water 10 Alternating current 11 Equipotentiality 12 Fuse 13 Foot...

Страница 16: ...nd or saphenous vein grafts See Information For Use manuals for specific and complete prescribing information 2006 2009 MEDRAD INC All rights reserved All information drawings and ideas herein constit...

Отзывы: