The Console will load the pump and process
information from a bar code located on the
pump. After the Console successfully identifies
the catheter, the status panel will display the
model of catheter in use (Figure 9).
The Console will automatically prime the pump.
5
Place the foot pedal within easy access of the
physician. Choose a location that will minimize
accidental activation.
The status panel displays PRIME indicating that
the initial pump prime was successful
(Figure 10).
Catheter priming time is determined by the
catheter model and is automatically set by the
Console.
1
Prime the catheter by completely submerging
the tip in heparinized saline and pressing the
foot pedal (Figures 11 & 12).
9
Figure 11
Figure 10
7.3 Prime the Catheter
Figure 12
Figure 9