8.3.1
Cleaning of Lamps & Lamp Covers
A dust layer on the acrylic lamp covers and the UV tubes can reduce the irradiance. Therefore, it is
recommended to wipe these parts clean whenever needed, preferably with a dry soft microfibre cloth.
To remove an acrylic lamp
cover you first need to open
the black plastic caps of the
two quick-release fasteners.
Then unlock the fasteners by
turning these by 90° e.g. with
the aid of a coin. The other
side of the acrylic cover is held
by lugs which are inserted
into the cabin module. Pull
the open side of the acrylic
cover slightly forward, then
sideways to release the lugs
out of the cabin module, as
shown in the adjacent figure.
8.3.2
Cleaning underneath the Cabin
For cleaning the floor under the cabin, the floor plate of the cabin can be removed. To do so open the
cabin door, and unscrew the knob on the front side below the floor plate. Then lift the foot plate, with
the rear side first, upwards from the base frame, and out of the cabin.
8.4
Electromagnetic Compatibility (EMC)
This device has been tested and found to comply with the EMC limits of the international standard EN
60601-1-2. These limits are designed to provide reasonable protection against interference in a typical
medical installation. The System can radiate radio frequency energy if not installed in accordance with
the instructions, and may cause harmful interference to other devices in the vicinity. There is no
guarantee that interference will not occur in a particular installation. If the System does cause
interference with other devices, which can be determined by turning the System off and on, the user is
encouraged to try to correct the interference by one or more of the following measures:
a)
Re-orient or relocate the receiving device.
b)
Increase the separation between the equipment.
c)
Connect the system to an outlet socket on a circuit different from that to which the other
device(s) is connected.
Unlock quick-release fasteners
Remove acrylic lamp cover