3.2
Operation
3.2.1
Stand-By Mode
After switching on the device, the
touch screen shows the display shown
on the right. If your cabin is equipped
with only one lamp type, only the
output of the corresponding spectrum
will be displayed. The green bar to the
right of the mW/cm² value is for
orientation with respect to the
condition of the UV tubes. The more
the UV tubes lose power over time
due to wear, the more the green bar
decreases. At very low output of the
UV tubes, the colour of the bar then
changes to red. In order to ensure an uninterrupted use of the cabin, it is advisable to replace the UV
tubes when the green bar is only half the original length. To proceed with the selection of the desired
therapy parameters, please press the C
ONTINUE
button.