8.2
Conditions for Transport, Storage & Operation
After each transport, check the device for damage. Do not operate a damaged unit, but
immediately inform our customer service department!
Transport and storage conditions must be adhered to.
Storage Conditions
(these values apply in the appropriate packaging for a period of 10 weeks)
Ambient temperature:
-30 °C to +55 °C
Humidity:
10% to 90%
Atmospheric pressure:
500 hPa to 1060 hPa
Operation Conditions
Ambient temperature:
+5 °C to +35 °C
Humidity:
30% to 75%
Atmospheric pressure:
700 hPa to 1060 hPa
8.3
Cleaning & Disinfection
The device and in particular all components made of plastic should not be cleaned with abrasive
cleaners, solvents, or alcohol-containing agents. For cleaning we recommend the use of a normal
household cleaner, such as dishwasher liquid. For surface disinfection we recommend "Mikrozid
sensitive liquid®" (Schülke & Mayr GmbH, Germany), which can also be obtained from MEDlight.
Clean or disinfect the equipment only when it is switched off. All surfaces should only be
wiped with a slightly damp cloth to avoid liquid entering the device.