P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 73
-
interval and then press
Set
.
5.
Roll the trackball and press
Set
to begin a new time measurement. You can
measure a maximum of four groups of data. The outcome will be displayed in
the measurement result window, as shown below.
6.
Press
Measure
to finish and exit.
Time1
Figure 6-15 Time Measurement
Slope
To measure slope:
1.
Press
Measure
to activate a measurement cursor “+”.
2.
Roll the trackball to highlight
Slope
and press
Set
and a big “+” is displayed.
3.
Roll the trackball and press
Set
to anchor the start point, and displays a big “+”.
4.
Roll the trackball and press
Set
to anchor the end point.
5.
Roll the trackball and press
Set
to begin a new slope measurement. You can
measure a maximum of four groups of data. The outcome will be displayed in
the measurement result window, as shown below.
6.
Press
Measure
to finish and exit.
Time
Depth
Slope1
Figure 6-16 Slope Measurement and the Results
Heart Rate
To measure heart rate:
1.
In the
B/M mode
, roll the trackball to change the position of the M Mark and
press
Set
to obtain a satisfying electrocardiogram, and then freeze it.
2.
In the
M mode
, freeze the desired image.
Measure the distance between two peaks of cardiac cycles with the time
measurement method.
1.
Press
Measure
to activate a measurement cursor “+”.
2.
Roll the trackball to highlight
Heart Rate
and press
Set
and a “+” is displayed.
3.
Roll the trackball to move the first measurement maker on the first peak systole