P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 147
-
Acoustic Output Reporting Table for Track3
System : P3V
Operating Mode:
B+M mode
Transducer: C611-2 Working Frequency: 6.5MHz
Index Label
MI
TIS
TIB
TIC
scan
non-scan
non-scan
A
aprt
≤1
A
aprt
>1
Global Maximum Index Value
0.78
0.11
0.0086
0.11
0.2726
Associated
Acoustic
Parameters
P
r.3
(MPa)
1.71
W
0
(mW)
7.7977
7.7977
7.7977
7.7977
Min of[W
.3
(z
1
),
I
TA.3
(z
1
)] (mW)
Z
1
(cm)
Z
bp
(cm)
Z
sp
(cm)
1.40
1.64
d
eq
(Z
sp
) (cm)
1.63
f
c
(MHz)
5.09
5.09
5.09
5.09
5.09
Dim of
A
aprt
X(cm)
0.672
0.672
0.672
0.672
Y (cm)
0.58
0.58
0.58
0.58
Other
Information
PD (usec)
0.26
PRF (Hz)
3911.5
P
r
@PII
max
(MPa)
2.18
d
eq
@PII
max
(cm)
1.63
Focal
Length
FL
x
(cm)
1.5
1.5
1.5
FL
y
(cm)
1.5
1.5
1.5
I
PA.3
@MI
max
(W/cm
2
)
108.17
Operating
Control
Conditions
Depth(mm)
29
29
29
29
29
Focus(mm)
25
25
25
25
25
Freq(MHz)
6.5
6.5
6.5
6.5
6.5