P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 124
-
11.2.1.
System Surface Cleaning
To clean the system surface:
1.
Power off the system and disconnect it from power supply.
2.
Use a clean gauze pad or lint-free cloth, moistened lightly with a mild detergent (Javel water),
to wipe the surface and the control panel.
3.
After cleaning, reconnect the system to power source.
CAUTION
Make sure the cleaning solution does not seep into the control panel or any other
openings.
NOTE:
1. Take particular care when cleaning the areas near trackball and the slide controls.
2. Make sure they are free of gel and any other visible residue.
3. Use a soft dry cloth without chemicals for cleaning, because the surface of polarizer
is very soft and easily scratched.
11.2.2. Probe and Probe Holder Cleaning and Disinfection
To clean probe holder:
1.
Disassemble the probe holder by uninstalling the two screws.
2.
Wash the holder with flowing water, using a mild detergent (Javel water).
3.
After cleaning and drying, assemble the probe holder to the main unit.
Every time before use, layer of medical ultrasound coupling gel should be applied evenly on the
area of the acoustical window of the probe. Be careful not to generate any air bubble.
To clean the probes:
1.
Disconnect the probe from the system.
2.
Wipe off all the gel gently with a soft cloth.
3.
Rinse the probe with enough distilled water to remove all visible soap residues.
4.
Air dry or dry with a soft cloth.
NOTE
:
The single-use sheath should be used on V563-2 probe. Before cleaning the
probe, remove the sheath gently and discard it. Put on a new single-use sheath before
using the probe.