P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 87
-
WARNING
1. Prior to each puncture, calibrate the needle guide line.
2. If the needle position is not the same as the needle guide line position, do not perform
the puncture.
To Display or To Hide the Needle Guide Line
Highlight
Display
in the needle guide menu, and press
Set
repeatedly to display or to hide the
needle guide line.
To Adjust the Needle Guide Line Position
Highlight
Position
in the needle guide menu, and press
Set
or
Back
to adjust the guide line
position horizontally.
To Select the Angle of Needle Guide Line
Needle guide line has been verified when the device is produced. The value is saved in Factory
data. But after a period of use, the needle guide line needs to be adjusted since the real value may
be changed. For instance, highlight
Angle
in the needle guide menu and press
Set
to adjust, and
the system displays the angle.
6.9.2. Calibrating Guide Line (Performing Phantom Cal)
1.
Put the assembled puncture kit, put the probe in to water phantom, and perform the water
scan;
2.
Adjust the position and angle of needle to comply with the guide line as shown below;
3.
Select
Verify
to save the verified value.
To verify the needle guide line:
Move the needle guide line horizontally
Highlight
Position
, press
Set
to increase the value and press
Back
to decrease the value,
and the value is displayed in the menu option.
To adjust the angel of needle guide line:
Enter
Angle
option to adjust the angle, press
Set
to increase the value and press
Back
to
decrease the value, and the value is displayed in the menu option.
To save the verified value:
After verifying the position and the angle, highlight
Verify
and press
Set
, and the system
will save the verified value. After restarting the system, the verified value is activated.
To save the factory data: