P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 113
-
The operations in these methods are similar to those in the single plane ellipse method. Please
refer to the corresponding B mode generic measure method for details, and you can use the
prompt instruction to help you.
CO calculation is as below.
Measurement and input items:
Measure LV;
Key in: HR
To calculate CO:
1.
In the B-cardiac measurement menu, roll the trackball to highlight
Input
. Then select the
secondary menu
Heart Rate
and press
Set
to display an
HR input
dialog box, as shown
below.
Figure 9-6 HR Input Dialog Box
2.
Input a suitable value in the HR (bpm) box.
3.
Roll the trackball to highlight
OK
and press
Set
. After measuring LV, CO will be
displayed in the measured result window.
MVCF calculation is as below.
Measurement and input items:
Measure: LV;
Key in: LVET
To calculate MVCF:
1.
Move the cursor to
Input
. Then select the secondary menu
LVET
and press
Set
to
display an
ET input
dialog box, as shown below.