5-8
P11 EXP
Portable Digital Color Doppler Ultrasound System
F4
Figure 5.4:
Biopsy Screen
And one biopsy line displays on the screen, as shown in the following figure.
You can adjust the biopsy offset and biopsy angle if necessary. It the
BIOPSY
is set to
OFF, no submenu displays.
✖
Remarks:
•
The range of the biopsy offset and angle vary with the probe.
•
You should calibrate the biopsy line each time before performing
biopsy. If the position of the biopsy needle is different with that of
the biopsy line, the biopsy cannot be performed.
•
The biopsy offset and angle restore to the factory default settings
after the system is restarted.
5.3.13
Sector Width and Position
User can reduce the imaging area to the region of interest by reducing the sector width and
move the sector position.
SEC.WIDTH
• Press F3 to select SEC.WIDTH.
• turn the function knob to change the sector width.
SEC.POS
• press
to select
SEC.POS
.
• turn the function knob to change the sector position.
BIOPSY OFFSET 12.2
BIOPSY ANGLE 42
Содержание P11 EXP
Страница 1: ...User Manual P11 EXP Ultrasound System Version 1 1 ...
Страница 4: ...P11 EXP Portable Digital Color Doppler Ultrasound System 0 2 ...
Страница 56: ...P11 EXP Portable Digital Color Doppler Ultrasound System 4 4 6 5 Annotation Edit Figure 4 11 Annotation edit ...
Страница 80: ...P11 EXP Portable Digital Color Doppler Ultrasound System 5 16 ...
Страница 102: ...8 8 P11 EXP Portable Digital Color Doppler Ultrasound System ...
Страница 118: ...P11 EXP Portable Digital Color Doppler Ultrasound System 10 10 ...
Страница 126: ...P11 EXP Portable Digital Color Doppler Ultrasound System 12 6 ...
Страница 136: ...P11 EXP Portable Digital Color Doppler Ultrasound System 13 ...
Страница 146: ...P11 EXP Portable Digital Color Doppler Ultrasound System A 6 ...
Страница 148: ...B 2 P11 EXP Portable Digital Color Doppler Ultrasound System ...