P11 EXP
Portable Digital Color Doppler Ultrasound System
A-2
Electromagnetic emissions
The EQUIPMENT is intended for use in the electromagnetic environment specified
below. The customer or the user of the EQUIPMENT should assure that it is used in
such an environment.
Emission test Compliance Electromagnetic environment guidance
The EQUIPMENT uses RF energy only for its
RF emissions
CISPR 11
RF emissions
Group 1
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
CISPR 11
Harmonic
Class B
The EQUIPMENT is suitable for use in all
establishments, including domestic
establishments and those directly connected to
emissions
IEC 61000-3-2
Voltage
fluctuations /
flicker emissions
IEC 61000-3-3
Class A
Complied
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Содержание P11 EXP
Страница 1: ...User Manual P11 EXP Ultrasound System Version 1 1 ...
Страница 4: ...P11 EXP Portable Digital Color Doppler Ultrasound System 0 2 ...
Страница 56: ...P11 EXP Portable Digital Color Doppler Ultrasound System 4 4 6 5 Annotation Edit Figure 4 11 Annotation edit ...
Страница 80: ...P11 EXP Portable Digital Color Doppler Ultrasound System 5 16 ...
Страница 102: ...8 8 P11 EXP Portable Digital Color Doppler Ultrasound System ...
Страница 118: ...P11 EXP Portable Digital Color Doppler Ultrasound System 10 10 ...
Страница 126: ...P11 EXP Portable Digital Color Doppler Ultrasound System 12 6 ...
Страница 136: ...P11 EXP Portable Digital Color Doppler Ultrasound System 13 ...
Страница 146: ...P11 EXP Portable Digital Color Doppler Ultrasound System A 6 ...
Страница 148: ...B 2 P11 EXP Portable Digital Color Doppler Ultrasound System ...