• Cracks on the probe connector or any other kinds of visible damage.
• Deformed pins or broken pins exist inside the probe connectors.
Body surface probe usages:
Inspect the probe before each use. In case the following damage is found, stop using the
probe immediately.
1. Wear medical sterile gloves.
2. Orient the probe.
3. Perform a scan.
14.1.3
Cleaning the Probe
• To avoid probe damages, do not bump the probe on hard surfaces during the cleaning.
• To avoid the electric shock, disconnect the probe from the ultrasound system before
the cleaning.
• To avoid the potential disease transmission, you should wear medical sterile gloves
and protective goggle during the cleaning.
You should clean the probe after each use. Perform the following steps to clean the probe.
1. Disconnect the probe from the ultrasound system, and remove the probe sheath from
the probe.
2. Use a lint-free soft cloth dampened with mild soapy water to wipe the probe.
If the probe carries blood, body fluid or dried stains, you should wipe it with enzymatic
cleaner.
✖
Remark:
Follow the
manufacturer’s
instructions to make up the enzymatic cleaner.
If there are stains on surface or grooves of the probe, you should use a brush
dampened with portable water to remove them.
3. Rinse the surfaces of the probe with the fresh running water to remove the residual
cleaner completely.
4. Dry the probe with a lint-free soft dry cloth.
14.1.4
Disinfecting or Sterilizing the Probe
• Use the liquid disinfectant/sterilant that meets the local laws and regulations.
• If you use the recommended disinfectant/sterilant, ensure its level and the probe
soaking period meets the requirements. Otherwise, the probe could be damaged
and your warranty might be void. If you have other use or needs, ensure the level of
the make-up disinfectant/sterilant and the probe soaking period are suitable for the
intended clinical application.
• Do not use the expired disinfectant/sterilant.
• Store the probe in a sterile environment, and mention its expiry date before use.
• Do not disinfect/sterilize the probe through autoclaving or contact with ethylene oxide.
• Do not use thermal disinfection/sterilization. Temperatures higher than 66°C or 150◦ F
will damage the probe.
• Not allow any disinfectant/sterilant to be air-dried on the probe.
Содержание P11 EXP
Страница 1: ...User Manual P11 EXP Ultrasound System Version 1 1 ...
Страница 4: ...P11 EXP Portable Digital Color Doppler Ultrasound System 0 2 ...
Страница 56: ...P11 EXP Portable Digital Color Doppler Ultrasound System 4 4 6 5 Annotation Edit Figure 4 11 Annotation edit ...
Страница 80: ...P11 EXP Portable Digital Color Doppler Ultrasound System 5 16 ...
Страница 102: ...8 8 P11 EXP Portable Digital Color Doppler Ultrasound System ...
Страница 118: ...P11 EXP Portable Digital Color Doppler Ultrasound System 10 10 ...
Страница 126: ...P11 EXP Portable Digital Color Doppler Ultrasound System 12 6 ...
Страница 136: ...P11 EXP Portable Digital Color Doppler Ultrasound System 13 ...
Страница 146: ...P11 EXP Portable Digital Color Doppler Ultrasound System A 6 ...
Страница 148: ...B 2 P11 EXP Portable Digital Color Doppler Ultrasound System ...