M-142 Multifunction Calibrator
MEATEST, s.r.o.
98
User Manual v48
I/D (instrument identification)
*IDN?
This command returns the identification of the manufacturer, model, serial number and firmware revision.
The reply is formatted as follows:
MEATEST,M-142,412341,4.6
Operation complete
*OPC
<cr>
This command sets the OPC bit in the ESR (Event Status Register) when all pending operations are complete.
Operation complete?
*OPC?
<cr>
This command returns “1” to the output queue after all pending operations are complete.
Operation complete?
*OPC?
<cr>
This command returns “1” to the output queue after all pending operations are complete.
Wait-to-Continue command
*WAI
<cr>
Prevents the instrument from executing any further commands or queries until all previous remote commands
have been executed.
Reset
*RST
<cr>
This command resets the calibrator to its initial status.
Test operation
*TST?
<cr>
This command executes an internal self-test. Return the self-test result (“0” for pass or “1” for fail).
Status byte reading (IEEE488 only)
*STB?
<cr>
This query returns the Status Byte Register including the MSS bit.