M-142 Multifunction Calibrator
MEATEST, s.r.o.
48
User Manual v48
12.
Keyb.beep .... xx
ON/OFF
This parameter allows the acoustic indication of pressed buttons to be switched off or on. ON and OFF display
buttons can be used to switch the indication off or on. ON is set by the manufacturer.
This parameter does not control the acoustic indication of output voltages over 100 V and identification of errors.
13.
Keyb.volume .... xx
UP/DOWN
This parameter allows the volume of acoustic indication to be set. UP and DOWN display buttons allow to set
the value in the range of 00 to 15. The bigger the value, the louder sound. This parameter controls the volume of
keyboard beep (if switched on), indication of output voltages over 100 V and identification of errors when
controlling the calibrator.
14.
Brightness .... xx
UP/DOWN
This parameter sets the contrast of the display. UP and DOWN display buttons allow to set the value in the range
of 00 to 15.
15.
Rotary change .... xx ON/OFF
This parameter controls the range of functions of the potentiometer. If ON is set, the potentiometer can move the
cursor both to the left and right (
and
symbols) and up and down (
and
symbols). If OFF is set, the
potentiometer can move the cursor up and down only (
and
symbols).
16.
Switch polarity .... xx ON/OFF
This parameter controls the function of the built-in relay. If ON is set, the relay is closes before the test. If OFF is
set, the relay is opens before the test.
17.
Switch activity .... xx PASS/FAIL
This parameter controls the function of the built-in relay. If PASS is set, the relay is active (changes its status) if
the test results in PASS status. If FAIL is set, the relay is active (changes its status) if the test results in FAIL
status.
18.
Cal.code .... 00000
Entry of calibration code. Calibration code is a five-digit number, which must be entered to access the calibration
mode. If the calibration code is set to “00000”, this information is displayed in the Setup menu. Calibration code
can be changed. New calibration code can be directly entered using numeric keyboard and confirmed by pressing
ENTER. If non-zero calibration code is set, correct calibration code must be entered to access the calibration
mode. Non-zero calibration code is not displayed further on the display.
The purpose of the calibration code is to prevent unauthorized users from changing the calibration of the
instrument.