![Meatest M-142 Скачать руководство пользователя страница 96](http://html.mh-extra.com/html/meatest/m-142/m-142_operation-manuals_1758453096.webp)
M-142 Multifunction Calibrator
MEATEST, s.r.o.
96
User Manual v48
MEAS :CONF :VOLT
This command selects the function VOLT (range 0 .. 12V) of internal multimeter and switches it on.
Example: to set the multimeter to DC voltage 10V measurement mode:
MEAS :CONF :VOLT <cr>
MEAS :CONF :CURR
This command selects the function CURR (range 0 .. 25mA) of internal multimeter and switches it on.
Example: to set the multimeter to DC current 20mA measurement mode:
MEAS :CONF :CURR <cr>
MEAS :CONF :MVOLT
This command selects the function MVOLT (range 0 .. 2000mV) of internal multimeter and switches it on.
Example: to set the multimeter to DC voltage 2000mV measurement mode:
MEAS :CONF :MVOLT <cr>
MEAS :CONF :RES
This command selects the function RES (range 0 .. 2000
) of internal multimeter and switches it on.
Example: to set the multimeter to resistance measurement mode:
MEAS :CONF :RES <cr>
MEAS :CONF :FREQ
This command selects the function FREQ of internal multimeter and switches it on.
Example: to set the multimeter to frequency measurement mode:
MEAS :CONF :FREQ <cr>
MEAS :CONF :TEMP :RTD :TYPE (?) <CPD> { PT385 | PT392 }
This command selects the function RTD of internal multimeter, sets the type of resistance temperature sensor and
switches the multimeter on.
Example: to set the multimeter to resistance temperature sensor type PT385 measurement mode:
MEAS :CONF :TEMP : RTD :TYPE PT385 <cr>
If query is sent, M142 returns the set type of approximation table { PT385 | PT392 }.
MEAS :CONF :TEMP :RTD :NRESistance (?) <DNPD>
This command selects the function RTD of internal multimeter, sets the nominal resistance of the resistance
temperature sensor to 0
C and switches the multimeter on. 10
to 2k
can be set.