MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v48
53
The table lists recommended calibration points. Having selected the required calibration point using SELECT
display button, the following data are shown.
Display buttons have the following meaning:
WRITE
new calibration value is entered into the memory, old value is irreversibly lost
SKIP
current calibration step is skipped, old value is retained in the memory
EXIT
current calibration is terminated. After this button is pressed, the calibration memory
hold all data (old or new entered) and the calibrator returns to the calibration
menu. It is not necessary to calibrate all ranges; calibration of only selected ranges
is possible by skipping the ranges which do not need to be calibrated.
Moreover, the display shows the range which is being calibrated (RANGE), and the value to be set at the external
standard multimeter (VALUE).
Setting the new calibration data
Use
,
,
,
cursor buttons to set such main data on the display, when the output signal measured by external
standard multimeter reaches the required calibration point. When the standard output value is reached, press
“WRITE” to write new calibration value to the calibration memory. If you press “SKIP” button, the calibrator