MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v48
57
POWER DC function
DC current calibration
nominal value [A]
set limits [A]
range [A]
note
0.0
400 n
20 m
zero calibration
0.0
400 n
-20 m
zero calibration
19 m
2 u
20 m
slope calibration
-19m
2 u
-20 m
slope calibration
0.0
2 u
200 m
zero calibration
0.0
2 u
-200 m
zero calibration
190 m
20 u
200 m
slope calibration
-190 m
20 u
-200 m
slope calibration
0.0
50 u
2
zero calibration
0.0
50 u
-2
zero calibration
1.9
200 u
2
slope calibration
-1.9
200 u
-2
slope calibration
0.0
200 u
10
zero calibration
0.0
200 u
-10
zero calibration
10
1 m
10
slope calibration
-10
1 m
-10
slope calibration
Tab.10 DC POWER
POWER AC function
AC current calibration
nominal value [A]
set limits [A]
range [A] recommended frequency[Hz]
1.9 m
400 n
20 m
120
19 m
2 u
20 m
120
19 m
2 u
200 m
120
190 m
20 u
200 m
120
190 m
20 u
2
120
1.9
200 u
2
120
1.9
200 u
10
120
10
1 m
10
120
Tab.11 AC POWER
Other then recommended frequency can be used for calibration. Specification of the calibrator is valid when
recommended frequency is used.