![Meatest M-142 Скачать руководство пользователя страница 112](http://html.mh-extra.com/html/meatest/m-142/m-142_operation-manuals_1758453112.webp)
M-142 Multifunction Calibrator
MEATEST, s.r.o.
112
User Manual v48
Examples of tests
Examples of applications
Sourcing/simulating
Measurement
Application
Pt 100 temperature
sensor
10 V / 20 mA / f
Calibration of temperature evaluation units, setting of
temperature regulators
TC temperature sensor
10 V / 20 mA / f
Calibration of temperature evaluation units, setting of
temperature regulators
Frequency
10 V / 20 mA
Setting/calibration of energy meters
Resistance
10 V / 20 mA
Measurement of resistance bridges
Calibration of a single phase “electrical power/current loop” transducer
Function setup:
Calibrator
P function (voltage, current, power factor, frequency)
Multimeter
DCI
Adapter
Opt. 40
INPUT
OUTPUT
HU LU HI LI +I -I
DUT – power transducer
Opt. 40 adapter