11.3 Leakage Current
Leakage current is a primary indicator of electrical shock hazard to personnel making contact
with any exposed outer surface of the equipment. Each ANI Monitor V1 is carefully checked
at the factory to verify that leakage current meets the IEC60601-1 safety standards.
The ANI Monitor V1 should be checked routinely for leakage current at least once a
year.
Always have the leakage current checked after a saline or blood spill, or immediately after a
major surge in the house electrical system and every time the monitor case has been opened.
12
Specifications, warranty and software license agreement
12.1 Specifications
Power Requirements: 100-240 Volts, 50/60 Hz, 3A
Consummation: < 32 W
Electrical Safety: category 1 conforms to IEC 60601-1.
Battery: Li-ion, output: 12 V
+
/-5%, 3800mAH. Around 90 minutes of full using.
Power connexion: cable plug
Weight of the monitor (without any accessory): 6.99 lb.
Weight of the acquisition device (without the ANI sensors cable): 0.88 lb.
Dimensions of the monitor (inches): 10.43 width
x
9.72 height
x
3.13 thick
Screen size: 8.4’’, resolution 800
x
600
Dimensions of the acquisition device (inches): 6.18 length
x
4.06 width
x
2.7 thick
The efficacy and security of the products are warranted during the life time of the products (5
years from the manufacturing date).
Material included in the ANI Monitor V1:
-
Panel PC: Mdoloris Medical Systems reference: MN-ANI-V1
Manufacturer: IEI (product reference: AFL-08A-N270)
-
Acquisition device: Mdoloris Medical Systems reference: BA-ANI-V1
Manufacturer: RHEA Electronique