ANI Monitor V1 - Continouous analgesia monitoring system
MD/PRD/IN16.ANIV1 V.10 – 29 SEP 2021
45
Cleaning the monitor display:
clean the monitor display screen with a mild solution of detergent and warm water or a
commercial display screen cleaner, available through personal computer dealers. To avoid
scratching the screen, never use abrasive cleaners.
Disinfecting the monitor and
acquisition device
:
Use lint-free absorbent towels dampened with a 10% bleach solution, or a commercial
disinfectant. After cleaning, dry all areas except the monitor display screen with a lint-free
absorbent paper towel. Wipe the connector ends with alcohol and allow it to dry completely.
WARNING:
WHENEVER AN EVENT SUCH AS SPILLAGE OF BLOOD OR SOLUTIONS
OCCURS, RE-TEST LEAKAGE CURRENT BEFORE FURTHER USE. DO NOT MIX
DISINFECTING SOLUTIONS AS HAZARDOUS GASES MAY RESULT.
CAUTION:
do not autoclave the acquisition device or monitor. Autoclaving will seriously
damage both components.
Avoid liquid ingress with the connection cables. Contact of fluids with the connections cables
can interfere with the acquisition performance.
11.2 Maintenance
The ANI Monitor V1 is designed so that no periodic maintenance is required.
Suggested routine maintenance includes: periodic checking of cable integrity and checking of
leakage current (systematically after a saline or blood spill, or immediately after a major surge
in the house electrical system, or at least once a year).
Instructions on checking leakage current are included in the next chapter.