ANI Monitor V1 - Continouous analgesia monitoring system
MD/PRD/IN16.ANIV1 V.10 – 29 SEP 2021
18
3.4
Battery
The ANI Monitor V1 contains an internal battery. The internal battery of the ANI Monitor V1
allows the system to keep calculating the ANI while the patient is moved to another room or
in case of temporary interruption of electrical main supply. The internal battery is not
designed to withstand alone a long time monitoring. The battery gauge has ten graduations;
each symbolizes around 10 min of battery life. As soon as the electrical main supply is
interrupted, a message appears on the screen with a warning bell in order to push the user to
plug the monitor into an electrical main supply quickly. The ANI index is still calculated
during this event.
NOTE:
To ignite the monitor, it has to be plugged in first intention to prevent a sudden
electrical crash in case of the battery is not enough charged
.
Check the battery charge gauge displayed before unplugged: the battery could be not
sufficiently charged and interrupt the electrical supply of the ANI Monitor V1.
If the power plug has been disconnected whereas the battery was not sufficiently charged,
the battery could be unable to supply the energy the monitor needs, so it will brutally stopped.
Once you plug again the monitor and if the front switch (see figure 12 below) doesn’t reboot
it please use the green rear button (see figure 13 below) to power the monitor up again. The
power indicator becomes bright.
Figure 10
Figure 11