ANI Monitor V1 - Continouous analgesia monitoring system
MD/PRD/IN16.ANIV1 V.10 – 29 SEP 2021
43
10
Environnement
10.1
Shipping and Storage Environnement
The ANI Monitor V1 and its accessories can be stored or shipped within the following
environmental limits. Note that these limits apply to non-operational storage and shipping
situations.
Temperature: -20°C to +60°C
Humidity:
15% to 95% (non-condensing)
Pressure:
360 mmHg to 800 mmHg
Protect the monitor from sudden temperature changes that can lead to condensation within the
instrument. To minimize condensation, avoid moving the system between heated buildings
and outside storage. Once moved inside, allow the monitor to stabilize in the unopened
shipping container at the inside ambient temperature before unpacking and placing into
service. Before operating the system, wipe down all visible condensation and allow the
system to reach equilibrium at room temperature.
10.2
Operating Environnement
The ANI Monitor V1 is not designed for use in areas containing flammable gases or vapors.
WARNING:
EXPLOSION HAZARD: DO NOT USE THE ANI Monitor V1 IN A
FLAMMABLE ATMOSPHERE OR WHERE CONCENTRATIONS OF FLAMMABLE
ANESTHETICS MAY OCCUR.
The ANI Monitor V1 is designed to operate safely in a temperature from 5° to 40° C to an
altitude of 2000m, in a humidity range from 15% to 95% (non-condensing) and a pressure
from 360mm Hg to 800mm Hg.